An Assessment of Artifact Calibration Effectiveness for a Multifunction Calibrator
Review the catalog details below, then open the PDF in your browser or save it for later.
About this document
An Assessment of Artifact Calibration Effectiveness for a Multifunction Calibrator is available as a free PDF for Fluke 5700A. It is cataloged as a Reference Manual in the MetrologyManuals technical document library.
Catalog details include manufacturer Fluke, primary model 5700A, 12 pages.
Recognized text from the PDF
- An Assessment of Artifact Calibration Effectiveness for a Multifunction Calibrator Les Huntley Fluke Corporation Everett, WA 98206-9090 ABSTRACT Analysis of As Found, As Left and Production data for a large sample (>250) of Multifunction Calibrators from a...
- Fluke's was the introduction of the 5700A MultiFunction Calibrator.
- Analysis of As Found, As Left and Production data for a large sample (>250) of Multifunction Calibrators from a few months to more than six years old shows the instrument is generally very conservatively specified and that Artifact Calibration does return the...